
Example: Individual Phase Loss Test
The following command sequence will drop the output of phase A only for 10 cycles at 400
Hz.
/* Set up sequence
VOLT 115;FREQ 400 Set output voltage to 115 V at 400 Hz (all phases)
INST:COUP 0 Uncouple output phases
INST:NSEL 1 Select phase A
OUTP 1 Close output relay
OUTP:STAR:STAT 1 Enable start phase mode
OUTP:STAR:PHAS 90 Set start phase angle to 90 degrees on phase A
OUTP:DROP 0.025 Drop for 25 ms (10 cycles @ 400 Hz)
Output drop status query
The status of a drop in progress can be queried using the query format of the OUTP:DROP
command. This query will return a <Bool> indicating if the output time specified has expired
(0) or the drop is still in progress (1).
Command Syntax OUTPut:DROP <NRf>
Parameters <NRf> | none
*RST Value none
Examples OUTP:DROP 0.001
Query Syntax OUTP:DROP?
Returned Parameters 0 | 1
Related Commands OUTP:STAR:STAT OUTP:STAR:PHAS INST:COUP
INST:NSEL
Komentáře k této Příručce